Shanghai B&W Sensing Technology Co.,Ltd.
Newly Designed Vibration Table Matching Sensor
Extended low-temperature vibration table with matching sensors
Model (Standard)
| Model | 13187/13287 | 14187/14287 | 11259 | |
| Dimensions mm (Hex x Height) | 13x22.5/13.5x16 | 13x22.5/13.5x16 | 14x20(L square x H)) | |
| Feature Description | Replaces original 14108, 13108; compact design; standard sensor for calibration | |||
| Sensitivity (mV/g) | 50 | 100 | 10(±2%) | |
| Measurement Range (g peak) | ±100 | ±50 | ±500 | |
| Amplitude Linearity | ≤±1% | |||
| Broadband Resolution (g rms) | 0.0001 | 0.00005 | 0.0005 | |
| Frequency Range | ±5%(Hz~kHz) | —— | —— | 1~11 |
| ±10%(Hz~kHz) | 0.5~10 | 0.5~10 | 1~14 | |
| Installation Resonant Frequency(kHz) | ≥35 | ≥33 | ≥30 | |
| Transverse sensitivity | ≤5% | ≤3% | ||
| Power Supply (VDC) | 18~30 | |||
| Supply Current (mA) | 2~20 | |||
| Output Impedance (Ω) | ≤100 | |||
| DC Bias Voltage (V) | 8~12 | |||
Electrical Noise Power Spectral Density (μg/√Hz | 10Hz | 24 | 12 | 120 |
| 100Hz | 8 | 4 | 40 | |
| 1000Hz | 6 | 3 | 30 | |
| Install Insulation Resistance | —— | —— | —— | |
| Maximum Withstand Acceleration Value (g peak) | 800 | 400 | 3000 | |
| Maximum Withstand Impact Value (g peak) | 2000 | 1000 | 8000 | |
| Operating Temperature (°C) | -50~125 | |||
| Sealing Form | Laser Welding | |||
| Sensitive Materials | Ordinary Piezoelectric Ceramic/Shear | High-end Piezoelectric Ceramic/Shear | ||
| Case material | Stainless Steel | Titanium Alloy | ||
| Installation Method_Internal Threaded Screw Hole | M5 | Top: M5 (optional) Bottom: M5 | ||
| Weight (grams) | 12 | 15 | 20 | |
| Sensor Output Format | M5 Top/Side Output | M5 Side Output | ||
| Cable end connector type | M5—BNC(511xx12);M5—M5(511xx11) | |||
Common Cable Options (Cable Description/Model | * Cable end connectors M5—BNC: Φ3mm PVC cable (5111312); Φ3mm TPV cable (5112812) * M5—M5 connectors at both ends of the cable: Φ3mm PVC cable (5111311); Φ3mm TPV cable (5112811) | |||
| Standard Attachments | ①M5 mounting screws; ②Standard: 1-meter PVC coaxial cable; ③Factory calibration report | |||
| Other Options | * Model + "F": Optional 5m/10m waterproof integrated cable (IP68) Note: The high-frequency cutoff of the frequency response may vary due to the connecting cable. The specific high-frequency cutoff will be determined based on the sensor calibration report, except for the 11259 standard sensor | |||